Continuous Data Streaming
On power up, after the initial 60-second heater delay, the Mini-OX will automatically begin outputting the measured O2 concentration and sensor td as both an averaged and raw value.
The averaged values give a stable output with the amount of averaging user variable while the raw un-averaged values allow the user to detect sudden oxygen changes.
The averaged value is the measurement output on both the 4 ā 20mA and 0 ā 10VDC outputs. The sensor td value is the measure of the partial pressure of oxygen in the measurement gas. The O2 concentration (%) is the td value scaled by the stored calibration value.
To stop or restart the data streaming:
-
Connect the Mini-OX via the RS232 interface to the PC.
-
Type āSā (not case sensitive).
NOTE: Data streaming automatically ceases during calibration.